Measurements
R&S
®
RTE
363User Manual 1326.1032.02 ─ 20
Meas. type Description/Result
Eye amplitude Eye top - Eye base
Jitter (peak to peak) Average of the jitter for both crossing points.
Jitter = (σcrossing1 + σcrossing2) / 2
Jitter (6*σ) Jitter (6*σ) = Jitter * 6
Jitter (RMS) Quadratic mean of the jitter at both crossing points
8.2.6.2 Settings for eye diagram measurements
Access: "Meas" menu > "Setup" > "Eye" category
Eye diagram measurements are only available for sources in the time domain.
For eye measurements, no further settings are required.
8.2.7 Spectrum measurements
Spectrum analysis determines the frequencies of a given input signal over time. Vari-
ous measurements can then be performed based on the signal spectrum.
Automatic measurements