Power analysis (option R&S RTE-K31)
R&S
®
RTE
1018User Manual 1326.1032.02 ─ 20
Report comment
In this field you can write a comment that is displayed in the measurement report, for
example, a measurement description.
Execute
Starts the "Transient Response" measurement.
Remote command:
POWer:TRANsient:EXECute on page 1979
16.6.3 Output spectrum
"Output Spectrum" analysis measures the spectrum of the output voltage. The results
can be applied to see typical side effect problems of the SMPS application, such as
switching frequency components of internal SMPS.
Required probes:
●
Voltage probe
16.6.3.1 Output spectrum results
After executing the "Output Spectrum" measurement, the following windows are dis-
played:
The results of " Power Quality" measurements are provided in two ways:
●
Two diagrams that show the graphical presentation of:
– the voltage waveform
– the spectrum
●
The result box displays the positions of the measured peaks. The peaks are found
by an automatic peak search.
Note: If no results are found, check and correct the FFT settings.
To measure and display the output spectrum, the instrument uses the following mea-
surements and waveforms:
●
"M4" Math 4 to calculate the magnitude of the FFT for the voltage source values
●
"C 1" to "C 4": Cursor 1 to Cursor 4 to determine the positions of the peaks
The used resources are listed in the "Details" tab. See also: Chapter 16.2.3, "Details
tab", on page 966.
Output