Power analysis (option R&S RTE-K31)
R&S
®
RTE
1017User Manual 1326.1032.02 ─ 20
Edge trigger slope ← Trigger
Sets the edge type for the trigger event.
"Positive"
Selects the rising edge, that is a positive voltage change.
"Negative"
Selects the falling edge, that is a negative voltage change.
"Both"
Selects the rising as well as the falling edge.
Remote command:
POWer:TRANsient:TRGSlope on page 1981
Level ← Trigger
Sets the voltage or current level for the trigger event.
Remote command:
POWer:TRANsient:TRGLevel on page 1981
Tube tolerance
Specifies a tolerated error band for the signal level.
Remote command:
POWer:TRANsient:HYSTeresis on page 1979
Signal high
Sets the expected signal high-voltage value.
Remote command:
POWer:TRANsient:SIGHigh on page 1980
Signal low
Sets the expected signal low voltage value.
Remote command:
POWer:TRANsient:SIGLow on page 1980
Horizontal SMPS Frequency
Sets the SMPS switching frequency. Set the value according to your DUT.
Remote command:
POWer:TRANsient:FREQuency on page 1979
Derived scale
Shows the scale of the displayed results. This value is calculated as the inverse of the
"SMPS switching frequency"
Optimize scaling
Selects the scaling for the display of the results.
"Autoscale the signals"
Automatically selects the most appropriate scale for the display of the
results.
"Keep present settings"
The present display settings are not changed.
Remote command:
POWer:TRANsient:AUToscale on page 1979
Output