Remote control commands
R&S
®
RTE
1916User Manual 1326.1032.02 ─ 20
LOW
The level of a digital channel stays below the threshold, or the
logical expression for LOGic trigger source is false.
EITHer
The signal state remains unchanged.
*RST: HIGH
Usage: Asynchronous command
TRIGger<m>:PARallel:TIMeout:TIME <Time>
Defines the time limit for the timeout at which the instrument triggers.
Parameters:
<Time> Range: 100E-12 to 10000
Increment: 100E-9
*RST: 100E-9
Default unit: s
Usage: Asynchronous command
18.18.4.5 Data2Clock trigger
See also:
●
TRIGger<m>:PARallel:DATatoclock:CSOurce[:VALue] on page 1913
TRIGger<m>:PARallel:DATatoclock:CSOurce:EDGE....................................................... 1916
TRIGger<m>:PARallel:DATatoclock:STIMe..................................................................... 1916
TRIGger<m>:PARallel:DATatoclock:HTIMe.....................................................................1917
TRIGger<m>:PARallel:DATatoclock:CSOurce:EDGE <ClockEdge>
Sets the edge of the clock signal. The crossing of the clock edge and the logical
threshold defines the time reference point for the setup and hold time measurement.
Parameters:
<ClockEdge> POSitive | NEGative | EITHer
*RST: POSitive
Usage: Asynchronous command
TRIGger<m>:PARallel:DATatoclock:STIMe <SetupTime>
Sets the minimum time before the clock edge while data should be stable and not
change its state.
See also: "Setup time" on page 915
Mixed signal option (MSO, R&S