Remote control commands
R&S
®
RTE
1913User Manual 1326.1032.02 ─ 20
TRIGger<m>:PARallel:TYPE <Type>
Selects the trigger type to trigger on digital channels and parallel buses.
To trigger on analog channels or the external trigger input, use TRIGger<m>:TYPE.
Parameters:
<Type> EDGE | WIDTh | TIMeout | DATatoclock | STATe | PATTern |
SERPattern |
SERPattern = serial pattern trigger
*RST: EDGE
Usage: Asynchronous command
TRIGger<m>:PARallel:DATatoclock:CSOurce[:VALue] <ClockSource>
TRIGger<m>:PARallel:STATe:CSOurce:VALue <ClockSource>
TRIGger<m>:PARallel:SPATtern:CSOurce[:VALue] <ClockSource>
Selects the digital channel of the clock signal.
Parameters:
<ClockSource> D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 |
D13 | D14 | D15
*RST: D0
Usage: Asynchronous command
TRIGger<m>:PARallel:EDGE:EXPRession[:DEFine] <LogicalExpr>
TRIGger<m>:PARallel:WIDTh:EXPRession[:DEFine] <LogicalExpr>
TRIGger<m>:PARallel:TIMeout:EXPRession[:DEFine] <LogicalExpr>
TRIGger<m>:PARallel:STATe:EXPRession[:DEFine] <LogicalExpr>
TRIGger<m>:PARallel:PATTern:EXPRession[:DEFine] <LogicalExpr>
TRIGger<m>:PARallel:SPATtern:EXPRession[:DEFine] <LogicalExpr>
Defines a logical combination of several digital channels as trigger condition if
TRIGger<m>:SOURce[:SELect] is set to LOGIC.
Parameters:
<LogicalExpr> String with logical expression
Example:
TRIGger:PARallel:EDGE:EXPRession 'D1 and D2'
Usage: Asynchronous command
18.18.4.2 Edge trigger
See also:
●
TRIGger<m>:PARallel:EDGE:EXPRession[:DEFine] on page 1913
TRIGger<m>:PARallel:EDGE:SLOPe.............................................................................1914
Mixed signal option (MSO, R&S
RTE-B1)