Remote control commands
R&S
®
RTE
1179User Manual 1326.1032.02 ─ 20
Usage: Asynchronous command
TRIGger<m>:TIMeout:TIME <Time>
Defines the time limit for the timeout at which the instrument triggers.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
Parameters:
<Time> Range: 100E-12 to 10000
Increment: 100E-9
*RST: 100E-9
Default unit: s
Usage: Asynchronous command
18.9.8
Interval trigger
TRIGger<m>:INTerval:SLOPe....................................................................................... 1179
TRIGger<m>:INTerval:RANGe.......................................................................................1179
TRIGger<m>:INTerval:WIDTh........................................................................................1180
TRIGger<m>:INTerval:DELTa........................................................................................ 1180
TRIGger<m>:INTerval:SLOPe <Slope>
Sets the edge for the trigger. You can analyze the interval between positive edges or
between negative edges.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
Parameters:
<Slope> POSitive | NEGative
See Chapter 18.4.3, "Slope parameter", on page 1075.
*RST: POSitive
Usage: Asynchronous command
TRIGger<m>:INTerval:RANGe <RangeMode>
Defines the range of an interval in relation to the interval width specified using
TRIGger<m>:INTerval:WIDTh and TRIGger<m>:INTerval:DELTa.