Power analysis (option R&S RTE-K31)
R&S
®
RTE
981User Manual 1326.1032.02 ─ 20
16.4.1.1 Modulation analysis results
The results of "Modulation Analysis " measurements are provided in two ways:
The diagram shows the graphical presentation of:
the voltage or the current waveform
for "Type >Turn on" a track of the frequency and the duty cycle.
(Optional for "Type > Continuous" measurement) Two histograms display the den-
sity distribution of the measurement results in dependence of the frequency and
the positive duty cycle.
The result box displays the numeric measurement results.
To measure and display the power quality, the instrument uses the following measure-
ments and waveforms:
"Meas 7" to measure the positive duty cycle
"Meas 8" to measure the frequency
The used resources are listed in the "Details" tab. See also: Chapter 16.2.3, "Details
tab", on page 966.
Table 16-2: Statistic result parameters
Label Description
Current Currently measured value
+Peak Positive peak value (maximum)
-Peak Negative peak value (minimum)
μ (Avg) Average
RMS Root mean square
σ (S-dev) Standard deviation
Event count Number of measured pulses
Wave count Number of waveforms (acquisitions) the measurement is based on
"Modulation Analysis" is a statistical evaluation that is reset only if the measurement
setup is changed or you reset the statistics.
The following remote commands are used for handling the measurement results:
POWer:MODulation:RESult:ACTual? on page 1958
POWer:MODulation:RESult:AVG? on page 1958
POWer:MODulation:RESult:EVTCount? on page 1958