Power analysis (option R&S RTE-K31)
R&S
®
RTE
990User Manual 1326.1032.02 ─ 20
6. Select the correct channels for the "Current Source" and the "Voltage Source".
7. Select the "Settings" tab.
8. Select the "Source".
9. If "Average" is selected, enter the "Average count", that is the number of wave-
forms used for average calculation.
10. Set the reset condition for the average calculation:
If "Time" is selected, enter the "Reset time".
If "Waveforms" is selected, enter the "Reset count".
11. Select an "Optimize Scaling" option.
12. Tap "Execute".
On the screen you can see the measurement waveforms of the slew rate, the cur-
rent and the voltage. The result box with numeric measurement results is shown.
For details, see Chapter 16.4.3.1, "Slew rate results", on page 988.
16.4.3.3 Slew rate settings
In the "Channels" tab, you set the current source and the voltage source, see also:
Chapter 16.2.1, "Channels tab", on page 965.
In the "Settings" tab, you configure the slew rate measurement parameters and display
settings.
Source
Selects dV/dt or dI/dt as the source of the measurement.
Remote command:
POWer:SLEWrate:SOURce on page 1961