Power analysis (option R&S RTE-K31)
R&S
®
RTE
974User Manual 1326.1032.02 ─ 20
16.3.2.3 Inrush current settings
In the "Channels" tab, you set the current source, see also: Chapter 16.2.1, "Channels
tab", on page 965.
In the "Settings" tab, you configure the inrush current measurement parameters and
display settings.
Maximum current
Sets the maximum expected current for the vertical scale. Set the value according to
your signal to avoid saturation.
Remote command:
POWer:INRush:MAXCurrent on page 1954
Trigger current value (T)
Sets the current value for the trigger. The measurement starts after the signal of the
DUT reaches this current value.
Remote command:
POWer:INRush:TRIGger on page 1954
Gate Configuration
In this table, you can configure different gates (time periods). You can configure up to
five different gates. The time periods of the defined gates may overlap.
To add a gate press "Insert" or"Append". To remove a gate press "Remove".
Input