Remote control commands
R&S
®
RTE
1778User Manual 1326.1032.02 ─ 20
BUS<m>:MDIO:DATA:THReshold:LOW <ThresDatLow>
Defines the lower threshold level for the data signal.
Suffix:
<m>
.
1..4
Parameters:
<ThresDatLow> Range: -5 to 5
Increment: 0.1
*RST: 0.8
Default unit: V
BUS<m>:MDIO:PRESet
<Preset>
Selects the default threshold settings according to the Ethernet standard: 2.0 V and
0.8 V.
Suffix:
<m>
.
1..4
Parameters:
<Preset> DEFault | MANual
*RST: DEFault
BUS<m>:MDIO:COUPling <ThresCpl>
Overwrites the data thresholds with the clock thresholds.
Suffix:
<m>
.
1..4
Parameters:
<ThresCpl> ON | OFF
*RST: ON
18.17.16.2 Trigger
The trigger suffix <m> is always 1 and can be omitted. It selects the trigger event: Only
the A-trigger is available for triggering on serial buses.
To trigger on a serial bus, make sure that:
●
TRIGger<m>:SOURce[:SELect] is set to SBUS.
●
The sources of the serial bus are channel signals: use BUS<m>:...:SOURce
commands.
●
Decoding is enabled: BUS<m>[:STATe] is set to ON.
TRIGger<m>:MDIO:TYPE.............................................................................................1779
TRIGger<m>:MDIO:ST.................................................................................................1779
TRIGger<m>:MDIO:FRAMetype....................................................................................1780
Protocols