Remote control commands
R&S
®
RTE
1777User Manual 1326.1032.02 ─ 20
Parameters:
<SourceData> C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 |
D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15
Digital and analog channels cannot be used at the same time for
data and clock lines. For triggering on a serial bus, analog or
digital input channels are required.
See Chapter 18.4.2, "Waveform parameter", on page 1074
*RST: C1W1
BUS<m>:MDIO:CLOCk:THReshold:HIGH
<ThresClkHigh>
Defines the upper threshold level for the clock signal.
Suffix:
<m>
.
1..4
Parameters:
<ThresClkHigh> Range: -5 to 5
Increment: 0.1
*RST: 2
Default unit: V
BUS<m>:MDIO:CLOCk:THReshold:LOW <ThresClkLow>
Defines the lower threshold level for the clock signal.
Suffix:
<m>
.
1..4
Parameters:
<ThresClkLow> Range: -5 to 5
Increment: 0.1
*RST: 0.8
Default unit: V
BUS<m>:MDIO:DATA:THReshold:HIGH <ThresDatHigh>
Defines the upper threshold level for the data signal.
Suffix:
<m>
.
1..4
Parameters:
<ThresDatHigh> Range: -5 to 5
Increment: 0.1
*RST: 2
Default unit: V
Protocols