Triggers
R&S
®
RTE
242User Manual 1326.1032.02 ─ 20
● Timeout................................................................................................................. 250
● Interval.................................................................................................................. 251
● Slew rate............................................................................................................... 252
● Data2Clock trigger................................................................................................ 254
● State......................................................................................................................255
● Pattern...................................................................................................................256
● Serial pattern.........................................................................................................258
● TV/Video trigger.................................................................................................... 259
● Line trigger............................................................................................................ 263
● Triggering on serial buses.....................................................................................264
● Triggering on parallel buses and digital channels................................................. 264
6.3.1 Basic trigger settings
Access: [TRIGGER] > "Setup" tab
The basic trigger settings are the trigger source and the trigger type, including the trig-
ger level. These settings are specific for each condition in a trigger sequence. To set
the trigger level automatically, use "Find level".
Depending on the trigger type, additional settings are available. They are explained in
the trigger-type specific sections.
Source
Selects the source of the trigger signal for the current trigger condition. The trigger
source works even if it is not displayed in a diagram. It must be synchronized to the
signal to be displayed and analyzed.
The trigger source can be:
●
Channel 1...4: An analog input channel
●
Extern: External analog signal connected to the external trigger input
The external trigger source is supported for the "A only" sequence. It is not availa-
ble if a longer trigger sequence is selected, or if qualification is enabled.
●
Line: The instrument generates the trigger from the AC power input and synchroni-
zes the signal to the AC power frequency. Use this source if you want to analyze
Trigger types