Protocol analysis
R&S
®
RTE
611User Manual 1326.1032.02 ─ 20
Remote commands:
●
BUS<m>:LIN:FRAMe<n>:SDATa? on page 1539
●
BUS<m>:LIN:FRAMe<n>:SDEXport? on page 1540
13.6.5 LIN decode results
When the configuration of the serial bus is complete, the signal can be decoded:
1. In the "Protocol" dialog > "Decode" tab, enable "Decode".
2. In the "Protocol" dialog > "Display" tab, select additional result display settings:
"Show decode table" and "Show binary signals".
For a description of the display settings, see also Chapter 13.1.3, "Display",
on page 515.
The instrument captures and decodes the signal according to the standard definition
and the configuration settings.
The color-coding of the various protocol sections and errors simplifies the interpretation
of the visual display. The decode information condenses or expands, depending on the
horizontal scale. Various data formats are available to show the result values.
The endianness setting ("Transfer") is a trigger setting and not considered for decod-
ing. The binary results of data bytes are displayed MSB first.
LIN (option R&S
RTE-K3)