Remote control commands
R&S
®
RTE
1540User Manual 1326.1032.02 ─ 20
BUS<m>:LIN:FRAMe<n>:SDEXport?
Returns the symbolic data of the frame in export format.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<SymbolicData>
Usage: Query only
BUS<m>:LIN:FRAMe<n>:SYMBol?
Returns the symbolic label of the specified frame if the label list is enabled.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the number of the frame in the current acquisition, 1...n.
Return values:
<Label> String with symbolic name of the identifier
Example:
BUS:LIN:FRAMe2:SYMBol?
Response: Temperature
Usage: Query only
BUS<m>:LIN:FRAMe<n>:VERSion?
Returns the version of the LIN standard for the specified frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<FrameVersion> V1X | V2X | UNK
UNK: Unknown
*RST: UNK
Usage: Query only
Protocols