Remote control commands
R&S
®
RTE
1247User Manual 1326.1032.02 ─ 20
Suffix:
<m>
.
1 | 2
Selects the cursor set.
Return values:
<Delta> Range: -100E+24 to 100E+24
*RST: 0
Usage: Query only
Asynchronous command
CURSor<m>:YDELta:SLOPe <DeltaSlope>
Returns the inverse value of the voltage difference - the reciprocal of the vertical dis-
tance of two horizontal cursor lines: 1/
Δ
V.
Suffix:
<m>
.
1 | 2
Selects the cursor set.
Parameters:
<DeltaSlope> Range: -100E+24 to 100E+24
Increment: 0
*RST: 0
Usage: Asynchronous command
18.11.3 Peak search using cursors
CURSor<m>:FFT:SETCenter........................................................................................ 1247
CURSor<m>:FFT:TOCenter.......................................................................................... 1247
CURSor<m>:MAXimum[:PEAK].....................................................................................1248
CURSor<m>:MAXimum:LEFT.......................................................................................1248
CURSor<m>:MAXimum:RIGHt......................................................................................1248
CURSor<m>:MAXimum:NEXT...................................................................................... 1248
CURSor<m>:THReshold...............................................................................................1248
CURSor<m>:PEXCursion............................................................................................. 1249
CURSor<m>:FFT:SETCenter
Sets the center frequency to the frequency value that is measured at cursor line c1.
Suffix:
<m>
.
The suffix is irrelevant.
Usage: Event
Asynchronous command
CURSor<m>:FFT:TOCenter
Sets the vertical cursor line c1 to the center frequency.
Cursor measurements