Remote control commands
R&S
®
RTE
1177User Manual 1326.1032.02 ─ 20
OUTSide
Triggers if the signal stays above the upper level or below the
lower level for a specified time. The time is defined using the
TRIGger<m>:WINDow:TIME command.
*RST: ENTer
Usage: Asynchronous command
TRIGger<m>:WINDow:TIME <TimeRangeMode>
Defines the limit of the window in relation to the time specified using TRIGger<m>:
WINDow:WIDTh and TRIGger<m>:WINDow:DELTa. Time conditioning is available for
TRIGger<m>:WINDow:RANGe= "WITHin" and "OUTSide".
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
Parameters:
<TimeRangeMode> WITHin | OUTSide | SHORter | LONGer
WITHin
Triggers if the signal stays inside or outside the vertical window
limits at least for the time Width - Delta and for Width + Delta at
the most.
OUTSide
"Outside" is the opposite definition of "Within". The instrument
triggers if the signal stays inside or outside the vertical window
limits for a time shorter than Width - Delta or longer than Width +
Delta.
SHORter
Triggers if the signal crosses vertical limits before the specified
"Width" time is reached.
LONGer
Triggers if the signal crosses vertical limits before the specified
"Width" time is reached.
*RST:
WITHin
Usage: Asynchronous command
TRIGger<m>:WINDow:WIDTh <Width>
For the ranges "Within" and "Outside" (defined using TRIGger<m>:WINDow:RANGe),
the width defines the center of a time range which is defined by the limits "±Delta" (see
TRIGger<m>:WINDow:DELTa on page 1178).
For the ranges "Shorter" and "Longer", it defines the maximum and minimum time
lapse, respectively.
Trigger