Remote control commands
R&S
®
RTE
1112User Manual 1326.1032.02 ─ 20
ACQuire:SRReal <RealSampleRate>
Sets the number of captured waveform points per second. It considers the samples of
the ADC, and the reduction of waveform points by decimation.
If interpolation is not active, the sample rate is the reciprocal value of the resolution
and thus also depends on the acquisition time and the record length.
If interpolation is active, the sample rate is limited to the ADC sample rate.
Parameters:
<RealSampleRate> Range: 5 Gsample/s
Increment: 1
*RST: 5 Gsample/s
Default unit: Sa/s
Usage: Asynchronous command
Firmware/software: FW 3.40
ACQuire:RESolution
<Resolution>
Indicates the time between two waveform points in the record.
Parameters:
<Resolution> A fine resolution with low values produces a more precise wave-
form record.
Range: 1E-15 to 0.5
Increment: 10E-12
*RST: 100E-12
Default unit: s
Usage: Asynchronous command
ACQuire:POINts[:VALue] <RecordLength>
Indicates the record length, the number of recorded waveform points that build the
waveform across the acquisition time. [:VALue] can be omitted.
Parameters:
<RecordLength> Number of recorded waveform points.
Range: 1000 to 1000000000
Increment: 2
*RST: 1000
Default unit: pts
Usage: Asynchronous command
TIMebase:ROLL:ENABle <Mode>
Activates the automatic roll mode.
Acquisition and setup