Remote control commands
R&S
®
RTE
1111User Manual 1326.1032.02 ─ 20
Parameters:
<AutoAdjust> ON | OFF
*RST: ON
Usage: Asynchronous command
ACQuire:POINts:MAXimum <ReclgthLim>
Sets a limit for the record length to prevent very large records. This value only takes
effect if a constant resolution is selected with ACQuire:POINts:AUTO. If you increase
the time scale, the resolution remains constant and the record length increases until
the limit is reached. Further increase of the time scale changes the resolution and
keeps the record length limit.
Parameters:
<ReclgthLim> Range: 1000 to 800 MSa. The actual maximum can be
lower depending on the installed options, number of
active channels, measurements and math wave-
forms.
Increment: 2
*RST: 10E+6
Default unit: pts
Usage: Asynchronous command
ACQuire:POINts:ARATe?
Retrieves the sample rate of the ADC, that is the number of points that are sampled by
the ADC in one second.
Return values:
<ADCSampleRate> Range: 5 Gsample/s
Default unit: Sa/s
Usage: Query only
Asynchronous command
ACQuire:SRATe <SampleRate>
Defines the sample rate, that is the number of recorded waveform samples per sec-
ond.
Parameters:
<SampleRate> Range: 5 Gsample/s
Increment: 1
*RST: 5 Gsample/s
Default unit: Sa/s
Usage: Asynchronous command
Acquisition and setup