Remote control commands
R&S
®
RTE
1136User Manual 1326.1032.02 ─ 20
PROBe<m>:SETup:DMOFfset <DMOffset>
Sets the differential offset to compensate a DC voltage applied to the positive and the
negative input terminal.
Suffix:
<m>
.
1..4
Parameters:
<DMOffset> In "DM" probe mode (DMODe), this offset is used as channel off-
set and considered automatically for correction.
Range: -100E+24 to 100E+24
Increment: 1E-3
*RST: 0
Default unit: V
Usage: Asynchronous command
PROBe<m>:SETup:CMOFfset <CMOffset>
Sets the common-mode offset. The setting is only available for differential probes.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<CMOffset> Range: -100E+24 to 100E+24
Increment: 1E-3
*RST: 0
Default unit: V
Usage: Asynchronous command
PROBe<m>:SETup:NOFFset <NOffset>
Sets the negative offset to compensate a DC voltage applied to the negative input ter-
minal (Vp) referenced to ground.
Suffix:
<m>
.
1..4
Parameters:
<NOffset> In "N" probe mode (NMODe), this offset is used as channel offset
and considered automatically for correction.
Range: -100E+24 to 100E+24
Increment: 1E-3
*RST: 0
Default unit: V
Usage: Asynchronous command
Acquisition and setup