Remote control commands
R&S
®
RTE
1125User Manual 1326.1032.02 ─ 20
TRPRobe:SETup:BANDwidth?
PROBe<m>:SETup:BANDwidth?
Queries the bandwidth of the probe.
Suffix:
<m>
.
1..4
Selects the input channel.
Return values:
<Bandwidth> Range: 10000 to 20E+9
*RST: 1E+9
Default unit: Hz
Usage: Query only
Asynchronous command
TRPRobe:SETup:ATTenuation[:AUTO]?
PROBe<m>:SETup:ATTenuation[:AUTO]?
Queries the attenuation of the probe.
Suffix:
<m>
.
1..4
Selects the input channel.
Return values:
<PrbAttMdAuto> Range: 1E-3 to 1000
*RST: 1
Default unit: V/V
Usage: Query only
Asynchronous command
PROBe<m>:SETup:OFFSet:AZERo
Performs an automatic correction of the zero error. If the DUT is ground-referenced,
the AutoZero function can improve the measurement results.
See also: "Detect AutoZero, Use AutoZero" on page 195
Suffix:
<m>
.
1..4
Selects the input channel.
Usage: Event
Asynchronous command
PROBe<m>:SETup:OFFSet:USEautozero <AutoZeroOffset>
Includes the AutoZero offset in measurement results. The auto zero error is detected
with PROBe<m>:SETup:OFFSet:AZERo.