EasyManua.ls Logo

Rohde & Schwarz RTE - Page 1259

Rohde & Schwarz RTE
2078 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTE
1259User Manual 1326.1032.02 ─ 20
<MeasType>
parameter value
(remote control)
Measurement
(manual opera-
tion)
Description, result Suffix
<n>
DTOTrigger Delay to trigger Time between the trigger event and a following signal
slope. High accuracy even if the trigger event is outside
the acquisition data.
46
PROBemeter Trig. ProbeMeter DC voltage measured by the connected active R&S
probe
47
SLERising Slew rate on ris-
ing edge
Steepness of the rising edge: voltage difference between
the lower and higher reference level, divided by the rise
time.
48
SLEFalling Slew rate on fall-
ing edge
Steepness of the falling edge: voltage difference
between the higher and lower reference level, divided by
the fall time.
49
not used 50
MEASurement<m>:ENVSelect...................................................................................... 1259
MEASurement<m>:DETThreshold.................................................................................1260
MEASurement<m>:AMPTime:ALEVel............................................................................ 1260
MEASurement<m>:AMPTime:PFSLope......................................................................... 1260
MEASurement<m>:AMPTime:PSLope........................................................................... 1261
MEASurement<m>:AMPTime:DELay<n>:DIRection........................................................ 1261
MEASurement<m>:AMPTime:DELay<n>:ECOunt........................................................... 1262
MEASurement<m>:AMPTime:DELay<n>:SLOPe............................................................ 1262
MEASurement<m>:AMPTime:PTCount.......................................................................... 1263
MEASurement<m>:AMPTime:ESLope........................................................................... 1263
MEASurement<m>:AMPTime:CSLope........................................................................... 1263
MEASurement<m>:AMPTime:CLCK<n>:LSELect........................................................... 1264
MEASurement<m>:AMPTime:DATA<n>:LSELect............................................................1264
MEASurement<m>:AMPTime:DTOTrigger<n>:SLOPe.....................................................1264
MEASurement<m>:AMPTime:DTOTrigger<n>:LSELect................................................... 1265
MEASurement<m>:AMPTime:LCHeck<n>:VALid............................................................ 1265
MEASurement<m>:AMPTime:LCHeck<n>:LOWer:LIMit...................................................1265
MEASurement<m>:AMPTime:LCHeck<n>:UPPer:LIMit................................................... 1265
MEASurement<m>:AMPTime:LCHeck<n>:LOWer:MARGin............................................. 1266
MEASurement<m>:AMPTime:LCHeck<n>:UPPer:MARGin.............................................. 1266
MEASurement<m>:ENVSelect <EnvelopeCurve>
The command is only relevant for measurements on envelope waveforms. It selects
the envelope to be used for measurement.
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Automatic measurements