Remote control commands
R&S
®
RTE
1265User Manual 1326.1032.02 ─ 20
MEASurement<m>:AMPTime:DTOTrigger<n>:LSELect <DlyTrigLevSel>
Selects the reference level of the measurement source on which the delay is mea-
sured.
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
<n> 1..2
The suffix is irrelevant.
Parameters:
<DlyTrigLevSel> UPPer | MIDDle | LOWer
*RST: MIDDle
Usage: Asynchronous command
MEASurement<m>:AMPTime:LCHeck<n>:VALid <ValidRange>
Enables or disables limit checking for amplitude vs. time measurements in the speci-
fied measurement channel.
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
<n> 1..50
Number of the amplitude/time measurement, see Table 18-5.
Parameters:
<ValidRange> ILIMit | ULIMit | LLIMit | OLIMit
ILIMit
Inside (within) limit; between the upper and lower limit values
ULIMit
Upper limit; above the upper limit value
LLIMit
Lower limit; below the lower limit value
OLIMit
Outside limit; above the upper limit or below the lower limit val-
ues
*RST: ILIMit
Usage: Asynchronous command
MEASurement<m>:AMPTime:LCHeck<n>:LOWer:LIMit <LowerLimit>
MEASurement<m>:AMPTime:LCHeck<n>:UPPer:LIMit <UpperLimit>
Define the lower and upper limit for limit checking, respectiveley. The valid range is
defined using the MEASurement<m>:AMPTime:LCHeck<n>:VALid command.
Automatic measurements