Remote control commands
R&S
®
RTE
1650User Manual 1326.1032.02 ─ 20
Suffix:
<m>
.
1..4
Parameters:
<Polarity> NORMal | INVert
NORMal
Normal (non-inverted) data signal polarity
INVert
Inverted data signal polarity
*RST: NORMal
BUS<m>:ETHernet:THReshold:HIGH <ThresholdUpper>
Sets the upper threshold value for the signal digitization. If the signal value is higher
than the this threshold, the signal state is considered high.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdUpper> Range: 0 to 10
Increment: 0.01
*RST: 0.5
Default unit: V
BUS<m>:ETHernet:THReshold:LOW <ThresholdLower>
Sets the lower threshold value for the signal digitization. If the signal value is below this
threshold, the signal state is considered low.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdLower> Range: -10 to 0
Increment: 0.01
*RST: -0.5
Default unit: V
BUS<m>:ETHernet:PRESet <ThresholdPreset>
Sets the thresholds to predefined or individually definable voltage levels.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdPreset> T0 | T100 | TX0 | TX100 | MANual
T0
Sets the thresholds to the default values for 10BASE-T
(0 meters): upper threshold to 1.25 V, lower threshold to -1.25 V
Protocols