Remote control commands
R&S
®
RTE
1869User Manual 1326.1032.02 ─ 20
SEARch:TRIGger:SWIRe:ERRor:PARity <SearchName>,<ErrorParity>
SEARch:TRIGger:SWIRe:ERRor:PARity? <SearchName>
Enables searching for a parity error (even parity).
Parameters:
<ErrorParity> ON | OFF
*RST: ON
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:SWIRe:TIME:CONDition
<SearchName>,<TimeOperator>
SEARch:TRIGger:SWIRe:TIME:CONDition? <SearchName>
Set the condition for the data value of the time code. You can define an exact value or
a value range.
Parameters:
<TimeOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with the corresponding SEARch:TRIGger:SWIRe:
TIME:MIN command.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with SEARch:TRIGger:SWIRe:TIME:MIN and
SEARch:TRIGger:SWIRe:TIME:MAX.
*RST:
EQUal
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:SWIRe:TIME:MAX <SearchName>,<TimePatternMax>
SEARch:TRIGger:SWIRe:TIME:MAX? <SearchName>
Sets the the end value of a data type range for the time code if SEARch:TRIGger:
SWIRe:TIME:CONDition is set to INRange or OORange.
Parameters:
<TimePatternMax>
Parameters for setting and query:
<SearchName>
Protocols