EasyManua.ls Logo

Rohde & Schwarz RTE - Page 1175

Rohde & Schwarz RTE
2078 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTE
1175User Manual 1326.1032.02 ─ 20
SHORter
Triggers on runts shorter than the given "Runt width".
WITHin
Triggers if the runt length is inside a given time range. The range
is defined by "Runt width" and "±Delta".
OUTSide
Triggers if the runt length is outside a given time range. The
range is defined by "Runt width" and "±Delta".
*RST: ANY
Usage: Asynchronous command
TRIGger<m>:RUNT:WIDTh
<Width>
Defines the upper or lower voltage threshold. This command is not available if
TRIGger<m>:RUNT:RANGe is set to "ANY".
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
Parameters:
<Width> Range: 100E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
Usage: Asynchronous command
TRIGger<m>:RUNT:DELTa <WidthDelta>
Defines a range around the runt width specified using TRIGger<m>:RUNT:WIDTh.
This command is only available if TRIGger<m>:RUNT:RANGe is set to "WITHin" or
"OUTSide".
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
Parameters:
<WidthDelta> Range: 100E-12 to 864
Increment: 100E-9
*RST: 100E-12
Default unit: s
Usage: Asynchronous command
Trigger

Table of Contents

Other manuals for Rohde & Schwarz RTE

Related product manuals