Remote control commands
R&S
®
RTE
1638User Manual 1326.1032.02 ─ 20
TRIGger<m>:ARINc:LABel:MAX <LabelMax>
Sets the end value of a label pattern if TRIGger<m>:ARINc:LABel:CONDition is
set to INRange or OORange.
Parameters:
<LabelMax> Numeric or string pattern, see TRIGger<m>:ARINc:LABel:
MIN
TRIGger<m>:ARINc:MINGap:SELect <MinGapSelect>
Enables triggering when the minimum gap is out of range. You can set the minimum
gap with: TRIGger<m>:ARINc:MINGap:BITS.
Parameters:
<MinGapSelect> ON | OFF
*RST: ON
TRIGger<m>:ARINc:MINGap:BITS
<MinGapBits>
Sets the value for the minimum gap to be triggered on.
Parameters:
<MinGapBits> Range: 0 to 100
Increment: 1
*RST: 4
Default unit: bit
TRIGger<m>:ARINc:MAXGap:SELect <MaxGapSelect>
Enables triggering when the maximum gap is out of range. You can set the maximum
gap with: TRIGger<m>:ARINc:MAXGap:BITS.
Parameters:
<MaxGapSelect> ON | OFF
*RST: OFF
TRIGger<m>:ARINc:MAXGap:BITS <MaxGapBits>
Sets the value for the maximum gap to be triggered on.
Parameters:
<MaxGapBits> Range: 0 to 1000
Increment: 1
*RST: 100
Default unit: bit
Protocols