Remote control commands
R&S
®
RTE
1091User Manual 1326.1032.02 ─ 20
Usage: Asynchronous command
DISPlay:PERSistence:RESet
Resets the display, removing persistent values.
Usage: Event
Asynchronous command
DISPlay:INTensity <Intensity>
This value determines the strength of the waveform line in the diagram. Enter a per-
centage between 0 (not visible) and 100% (very strong).
The exact mapping of the cumulative value occurences according to the assigned color
table is guaranteed only if the intensity is set to 50% (default). All other intensity values
falsify the mapping but may improve the visibility of the signal.
See also: Chapter 4.4.3.2, "Changing waveform colors", on page 153.
Parameters:
<Intensity> Range: 0 to 100
Increment: 1
*RST: 50
Default unit: %
Usage: Asynchronous command
DISPlay:DIAGram:STYLe <Style>
Select the style in which the waveform is displayed.
Parameters:
<Style> VECTors | DOTS
VECTors
The individual data points are connected by a line.
DOTS
Only the individual data points are displayed.
*RST: VECTors
Usage: Asynchronous command
DISPlay:COLor:SIGNal<m>:COLor <Value>
Suffix:
<m>
.
1...118
Waveform number, see Chapter 18.4.1, "Waveform suffix",
on page 1072.
Parameters:
<Value>
Instrument setup