Remote control commands
R&S
®
RTE
1130User Manual 1326.1032.02 ─ 20
Return values:
<DiffMeasRes> Range: -100E+24 to 100E+24
Increment: 1E-3
*RST: 0
Default unit: V
Usage: Query only
Asynchronous command
Firmware/software: FW 2.25
TRPRobe:PMETer:RESults:NEGative?
PROBe<m>:PMETer:RESults:NEGative?
Returns the ProbeMeter measurement result of differential active R&S probes, the volt-
age that is measured between the negative signal socket and the ground.
Suffix:
<m>
.
1..4
Selects the input channel.
Return values:
<SgEndNegMeasRes>Range: -100E+24 to 100E+24
Increment: 1E-3
*RST: 0
Default unit: V
Usage: Query only
Asynchronous command
Firmware/software: FW 2.25
TRPRobe:PMETer:RESults:POSitive?
PROBe<m>:PMETer:RESults:POSitive?
Returns the ProbeMeter measurement result of differential active R&S probes, the volt-
age that is measured between the negative signal socket and the ground.
Suffix:
<m>
.
1..4
Selects the input channel.
Return values:
<Result> Range: -100E+24 to 100E+24
Increment: 1E-3
*RST: 0
Default unit: V
Usage: Query only
Asynchronous command
Firmware/software: FW 2.25