Remote control commands
R&S
®
RTE
1198User Manual 1326.1032.02 ─ 20
RANDom
Defines the holdoff as a random time limited by TRIGger<m>:
HOLDoff:MIN on page 1199 and TRIGger<m>:HOLDoff:MAX
on page 1199. For each acquisition cycle, the instrument selects
a new random holdoff time from the specified range.
AUTO
The holdoff time is calculated automatically based on the current
horizontal scale.
OFF
No holdoff
*RST: OFF
Usage: Asynchronous command
TRIGger<m>:HOLDoff:TIME <Time>
Defines the holdoff time period. The next trigger occurs only after this time has passed.
The setting is relevant if the holdoff mode is set to TIME.
See also:
●
TRIGger<m>:HOLDoff:MODE
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<Time> Range: 100E-9 to 10
Increment: 200E-6
*RST: 1E-3
Default unit: s
Example:
TRIGger:HOLDoff:MODE TIME
TRIGger:HOLDoff:TIME 1ms
The holdoff time is set to 1 ms.
Usage: Asynchronous command
TRIGger<m>:HOLDoff:EVENts <Events>
Defines the number of triggers to be skipped. The next trigger only occurs when this
number of events is reached. The setting is relevant if the holdoff mode is set to
EVENts.
See also:
●
TRIGger<m>:HOLDoff:MODE
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Trigger