Remote control commands
R&S
®
RTE
1216User Manual 1326.1032.02 ─ 20
Parameters for setting and query:
<DiagramName> Name of the diagram on which the zoom area is based.
<ZoomName> Name of the zoom diagram
Example: See Chapter 18.3.1.2, "Creating zoom diagrams", on page 1056
Usage: Asynchronous command
LAYout:ZOOM:HORZ:RELative:STARt <DiagramName>,
<ZoomName>,<RelativeStart>
LAYout:ZOOM:HORZ:RELative:STARt? <DiagramName>, <ZoomName>
Defines the lower limit of the zoom area on the x-axis.
Parameters:
<RelativeStart> Range: 0 to 100
Increment: 0.1
*RST: 0
Default unit: %
Parameters for setting and query:
<DiagramName> Name of the diagram on which the zoom area is based.
<ZoomName> Name of the zoom diagram
Usage: Asynchronous command
LAYout:ZOOM:HORZ:RELative:STOP <DiagramName>,
<ZoomName>,<RelativeStop>
LAYout:ZOOM:HORZ:RELative:STOP? <DiagramName>, <ZoomName>
Defines the upper limit of the zoom area on the x-axis.
Parameters:
<RelativeStop> Range: 0 to 100
Increment: 0.1
*RST: 100
Default unit: %
Parameters for setting and query:
<DiagramName> Name of the diagram on which the zoom area is based.
<ZoomName> Name of the zoom diagram
Usage: Asynchronous command
LAYout:ZOOM:VERTical:MODE <DiagramName>, <ZoomName>,<Mode>
LAYout:ZOOM:VERTical:MODE? <DiagramName>, <ZoomName>
Defines whether absolute or relative values are used to specify the y-axis values.
Since the zoom area refers to the active signal, relative values ensure that the zoom
area remains the same.