Remote control commands
R&S
®
RTE
1256User Manual 1326.1032.02 ─ 20
●
RELiability: reliability of the measurement result
●
RMS: RMS value of the long-term measurement results
●
STDDev: standard deviation of the long-term measurement results
For a detailed description of the results see "Selection of the measurement: MEASure-
ment<m>" on page 1249.
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Query parameters:
<MeasType> | Not relevant, omit it
Example: See "Simple frequency and amplitude measurement"
on page 1057.
See "Creating and reading histograms" on page 1058
Usage: Query only
MEASurement<m>:RESult:STARt? [<MeasType>]
MEASurement<m>:RESult:STOP? [<MeasType>]
Return the start and stop times of the selected measurement.
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Query parameters:
<MeasType> Not relevant, omit it
Usage: Query only
MEASurement<m>:RESult:COUNt?
Returns the number of result groups that are returned by MEASure
ment:RESult:ACTual? HAR.
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Query parameters:
<MeasType> HAR
The command is only relevant for harmonic search.
Return values:
Number Number of result groups