Remote control commands
R&S
®
RTE
1275User Manual 1326.1032.02 ─ 20
NOBorder
Label without border or lines.
UNDERLINE
The label is underlined.
FULL
The label is surrounded by a frame.
Parameters for setting and query:
<MeasType> PLISt
Only for peaklist measurements. For other measurement types,
this command returns an error.
MEASurement<m>:RESult:SHFRequency <MeasType>, <State>
MEASurement<m>:RESult:SHFRequency? <MeasType>
Includes the frequency of the detected peak in the diagram labels (if
MEASurement<m>:RESult:LABorder=ON).
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Parameters:
<State> ON | OFF
Parameters for setting and query:
<MeasType> PLISt
Only for peaklist measurements. For other measurement types,
this command returns an error.
Example: Display labels.
MEAS:RES:SHL PLIS,ON
Show frequency in labels.
MEAS:RES:SHFR PLIS,ON
Query the type of labels for peak lists.
MEAS:RES:SHFR? PLIS
//Result: ON
MEASurement<m>:RESult:SHLabels <MeasType>, <State>
MEASurement<m>:RESult:SHLabels? <MeasType>
Defines whether a description (label) is displayed for each detected peak in the spec-
trum diagram.
The layout of the label is defined by MEASurement<m>:RESult:LABorder.
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Automatic measurements