Remote control commands
R&S
®
RTE
1282User Manual 1326.1032.02 ─ 20
Parameters for setting and query:
<HistogramName>
Usage: Asynchronous command
LAYout:HISTogram:VERTical:RELative:STARt <HistogramName>,<RelativeStart>
LAYout:HISTogram:VERTical:RELative:STARt? <HistogramName>
Defines the vertical start value of the histogram.
Parameters:
<RelativeStart> Range: 0 to 100
Increment: 0.1
*RST: 0
Default unit: %
Parameters for setting and query:
<HistogramName>
Usage: Asynchronous command
LAYout:HISTogram:VERTical:RELative:STOP <HistogramName>,<RelativeStop>
LAYout:HISTogram:VERTical:RELative:STOP? <HistogramName>
Defines the vertical stop value of the histogram.
Parameters:
<RelativeStop> Range: 0 to 100
Increment: 0.1
*RST: 100
Default unit: %
Parameters for setting and query:
<HistogramName>
Usage: Asynchronous command
LAYout:HISTogram:RESet <HistogramName>
Resets the values to begin a new histogram.
Setting parameters:
<HistogramName>
Usage: Setting only
Asynchronous command
LAYout:HISTogram:REMove <Name>
Removes the specified histogram.
Setting parameters:
<Name>
Automatic measurements