Remote control commands
R&S
®
RTE
1321User Manual 1326.1032.02 ─ 20
Return values:
<FrameCoverage> Range: 0 to 100
*RST: 100
Default unit: %
Usage: Query only
Asynchronous command
CALCulate:MATH<m>:FFT:FRAMe:MAXCount <MaxFrameCount>
Restricts the maximum number of frames to be calculated. Due to the other parameter
settings, the required number of frames may become very high, thus slowing perfor-
mance. By restricting the number of frames, you can avoid performance loss without
changing the other parameters.
Suffix:
<m>
.
1..8
math waveform
Parameters:
<MaxFrameCount> Range: 1 to 10000
Increment: 10
*RST: 1000
Usage: Asynchronous command
CALCulate:MATH<m>:FFT:FRAMe:OFACtor <OverlapFactor>
Defines the minimum factor by which two neighboring frames overlap. If the required
number of frames to cover the input values allows for more overlap, the factor is
increased.
The higher the overlap factor, the more frames are used. This leads to more individual
results and improves detection of transient signal effects. However, it also extends the
duration of the calculation.
Suffix:
<m>
.
1..8
math waveform
Parameters:
<OverlapFactor> Range: 0 to 90
Increment: 1
*RST: 50
Default unit: %
Usage: Asynchronous command
CALCulate:MATH<m>:FFT:GATE:COUPling <GateRBWCoupling>
Defines the behaviour of the record length or RBW value in dependancy to the other
FFT parameters.
Spectrum analysis