Remote control commands
R&S
®
RTE
1445User Manual 1326.1032.02 ─ 20
SEARch:TRIGger:I2C:ADOR <SearchName>,<AddressOr>
SEARch:TRIGger:I2C:ADOR? <SearchName>
Enables the search for one to four address conditions.
●
SEARch:TRIGger:I2C:ADDO<m>:ENABle on page 1447
●
SEARch:TRIGger:I2C:ADDO<m>:ADRType on page 1447
●
SEARch:TRIGger:I2C:ADDO<m>[:VALue] on page 1448
●
SEARch:TRIGger:I2C:ADDO<m>:RWBit on page 1448
Parameters:
<AddressOr> ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:I2C:ADData <SearchName>,<AddressData>
SEARch:TRIGger:I2C:ADData? <SearchName>
Enables the search for a combination of address and data conditions.
Parameters:
<AddressData> ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:I2C:ACONdition <SearchName>,<AddressOperator>
SEARch:TRIGger:I2C:ACONdition? <SearchName>
Sets the operator to set a specific address or an address range.
Parameters:
<AddressOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These condition require one identifier pat-
tern to be set with SEARch:TRIGger:I2C:ADDRess
on page 1446.
INRange | OORange
In range / Out of range: Set the minimum and maximum value of
the range with SEARch:TRIGger:I2C:ADDRess on page 1446
and SEARch:TRIGger:I2C:ADDTo on page 1446.
*RST:
EQUal
Parameters for setting and query:
<SearchName>
Protocols