Remote control commands
R&S
®
RTE
1561User Manual 1326.1032.02 ─ 20
BUS<m>:FLXRay:PRDiff <PresetDiff>
Selects a default threshold voltage if BUS<m>:FLXRay:SRCType is set to
DIFFerential.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<PresetDiff> MV150 | MV200 | MV250 | MV300 | MANual
MV150 | MV200 | MV250 | MV300
±150 mV, ±200 mV, ±250 mV, ±300 mV respectively
MANual
Manual setting of user-defined values with BUS<m>:FLXRay:
THReshold<n> on page 1559.
*RST: MV150
BUS<m>:FLXRay:PRLogic
<PresetLogic>
Selects a default threshold voltage if BUS<m>:FLXRay:SRCType is set to LOGic.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<PresetLogic> V25 | V165 | V125 | V09 | V0 | MANual
V25 | V165 | V125 | V09 | V0
2.5 V (CMOS 5.0 V); 1.65 V (CMOS 3.5V), 1.25 V (CMOS 2.5V),
0.9 V (CMOS 1.8V), 0 V (ground)
MANual
Manual setting of user-defined values with BUS<m>:FLXRay:
THReshold<n> on page 1559.
*RST: V25
BUS<m>:FLXRay:POLarity <Polarity>
Selects the wire on which the bus signal is measured in case of SINGle source type.
The setting affects the digitization of the signal.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<Polarity> BPLus | BMINus
*RST: BPLus
Protocols