Remote control commands
R&S
®
RTE
1572User Manual 1326.1032.02 ─ 20
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<FrameState> OK | FSS | BSS | FES | INDicator | HCRCerror | CRCerr |
LENER | LENer | HCFCerror | INSufficient
OK: the frame is valid.
FSS: Frame Start Sequence after TSS is missing.
BSS: Byte Start Sequence is missing.
FES: error in the Frame End Sequence.
INDicator: Error in indicator bits.
HCRCerror: Header CRC is not valid.
CRCerr: Payload CRC is not valid.
LENER = LENer: Unexpected length of the frame.
HCFCerror: Header CRC error and frame CRC error
INSufficient: The frame is not completely contained in the acqui-
sition. The acquired part of the frame is valid.
*RST:
OK
Usage: Query only
BUS<m>:FLXRay:FRAMe<n>:STARt?
BUS<m>:FLXRay:FRAMe<n>:STOP?
Return the start time and stop time of the selected frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<FrameStop> Time
Range: -100E+24 to 100E+24
*RST: 0
Default unit: s
Usage: Query only
BUS<m>:FLXRay:FRAMe<n>:SYMBol?
Returns the symbolic label of the specified frame if the label list is enabled.
Suffix:
<m>
.
1..4
Selects the serial bus.