Remote control commands
R&S
®
RTE
1661User Manual 1326.1032.02 ─ 20
Return values:
<DestTranslation> String parameter
Usage: Query only
BUS<m>:ETHernet:WORD<n>:SSYMbol?
Returns the symbolic label (or translation) of the source address of the specified word,
if the label list is enabled.
Suffix:
<m>
.
1..4
<n> *
Return values:
<SrcTranslation> String parameter
Usage: Query only
BUS<m>:ETHernet:WORD<n>:BITRate?
Returns the primary bit rate.
Suffix:
<m>
.
1..4
<n> *
Return values:
<PrimaryBitRate> Range: 0 to 100000000000
Increment: 1
*RST: 0
Default unit: bps
Usage: Query only
BUS<m>:ETHernet:WORD<n>:BYTE<o>:VALue?
BYTE returns all data of up to 1982 bytes (not just the first 5 or 6 bytes). This is also
visible in the data table under "Show details".
Suffix:
<m>
.
1..4
<n> *
<o> *
Selects the byte number.
Return values:
<FrameByteValue> Range: 0 to 255
Increment: 1
*RST: 0