EasyManua.ls Logo

Rohde & Schwarz RTE - Page 1713

Rohde & Schwarz RTE
2078 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTE
1713User Manual 1326.1032.02 ─ 20
BUS<m>:SENT:FCOunt?
Returns the number of decoded frames of the acquisition.
Suffix:
<m>
.
1..4
Selects the serial data bus.
Return values:
<Count> Total number of decoded frames.
Range: 0 to 100000
Increment: 1
*RST: 0
Usage: Query only
BUS<m>:SENT:FRAMe<n>:STATus?
Returns the overall state of the selected frame.
Suffix:
<m>
.
1..4
Selects the serial data bus.
<n> *
Selects the frame.
Return values:
<FrameState> OK | SYNC | PULSe | CRC | IRFL | FORM | INSufficient
OK: the frame is valid.
SYNC: Synchronization error occurred.
PULse: Pulse error occurred.
CRC: Cyclic redundancy check failed.
IRFL: Irregular frame length error occurred.
FORM: Format error occurred.
INSufficient: The frame is not completely contained in the acqui-
sition. The acquired part of the frame is valid.
*RST: OK
Usage: Query only
BUS<m>:SENT:FRAMe<n>:STARt?
BUS<m>:SENT:FRAMe<n>:STOP?
Returns the start time and stop time of the selected frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Protocols

Table of Contents

Other manuals for Rohde & Schwarz RTE

Related product manuals