Remote control commands
R&S
®
RTE
1726User Manual 1326.1032.02 ─ 20
SEARch:TRIGger:SENT:PPERioderror <SearchName>,<PulsePeriodErr>
SEARch:TRIGger:SENT:PPERioderror? <SearchName>
Enables the search for pulse period errors.
To initially enable the search for an error event, set SEARch:TRIGger:SENT:ERRor
to ON.
Parameters:
<PulsePeriodErr> ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:SENT:IRFLength
<SearchName>,<IrregularFrmLen>
SEARch:TRIGger:SENT:IRFLength? <SearchName>
Enables the search for irregular frame length errors in transmission sequences.
This error type is relevant if pause pulse mode is set to constant frame length.
To initially enable the search for an error event, set SEARch:TRIGger:SENT:ERRor
to ON.
Parameters:
<IrregularFrmLen> ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:SENT:FORMerror <SearchName>,<FormError>
SEARch:TRIGger:SENT:FORMerror? <SearchName>
Enables the search for format errors in serial messages.
To initially enable the search for an error event, set SEARch:TRIGger:SENT:ERRor
to ON.
Parameters:
<FormError> ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:SENT:CRCerror <SearchName>,<CRCError>
SEARch:TRIGger:SENT:CRCerror? <SearchName>
Enables the search for errors in the Cyclic Redundancy Check.