Remote control commands
R&S
®
RTE
1976User Manual 1326.1032.02 ─ 20
POWer:RIPPle:RESult:NDCYcle:STDDev? [<VoltageMeasType>]
POWer:RIPPle:RESult:NDCYcle:WFMCount? [<VoltageMeasType>]
Return the specified statistic result for the negative duty cycle. The measurement
requires at least one complete period of a triggered signal.
[:ACTual]: current measurement result
AVG: average of the long-term measurement results
EVTCount: number of measurement results in the long-term measurement
NPEak: negative peak value of the long-term measurement results
PPEak: positive peak value of the long-term measurement results
RMS: RMS value of the long-term measurement results
STDDev: standard deviation of the long-term measurement results
Query parameters:
<VoltageMeasType> VOLTage | CURRent
*RST: VOLTage
Return values:
<Result> Statistic result
Usage: Query only
POWer:RIPPle:RESult:PDCYcle[:ACTual]? [<VoltageMeasType>]
POWer:RIPPle:RESult:PDCYcle:AVG? [<VoltageMeasType>]
POWer:RIPPle:RESult:PDCYcle:EVTCount? [<VoltageMeasType>]
POWer:RIPPle:RESult:PDCYcle:NPEak? [<VoltageMeasType>]
POWer:RIPPle:RESult:PDCYcle:PPEak? [<VoltageMeasType>]
POWer:RIPPle:RESult:PDCYcle:RMS? [<VoltageMeasType>]
POWer:RIPPle:RESult:PDCYcle:STDDev? [<VoltageMeasType>]
POWer:RIPPle:RESult:PDCYcle:WFMCount? [<VoltageMeasType>]
Return the specified statistic result for the positive duty cycle. The measurement
requires at least one complete period of a triggered signal.
[:ACTual]: current measurement result
AVG: average of the long-term measurement results
EVTCount: number of measurement results in the long-term measurement
NPEak: negative peak value of the long-term measurement results
PPEak: positive peak value of the long-term measurement results
RMS: RMS value of the long-term measurement results
STDDev: standard deviation of the long-term measurement results
Query parameters:
<VoltageMeasType> VOLTage | CURRent
*RST: VOLTage
Return values:
<Result> Statistic result