Remote control commands
R&S
®
RTE
1991User Manual 1326.1032.02 ─ 20
TRACe:REMote:MODE:FILE:STARtmode <StartMode>
Defines how the remote trace is started.
Parameters:
<StartMode> AUTO | EXPLicit
AUTO
Starts the remote trace immediately.
EXPLicit
Starts remote trace with TRACe:REMote:MODE:FILE:ENABle
ON
TRACe:REMote:MODE:FILE:STOPmode <StopMode>
Defines when the remote trace is stopped.
Parameters:
<StopMode> AUTO | EXPLicit | ERRor | BUFFerfull
AUTO
Ends the remote trace on device shutdown.
EXPLicit
Ends remote trace with TRACe:REMote:MODE:FILE:ENABle
OFF
ERRor
Ends remote trace when a SCPI error occurs.
BUFFerfull
Ends remote trace when the maximum file size is reached.
TRACe:REMote:MODE:FILE:ENABle <Enable>
Enables and disables the remote trace to file.
Parameters:
<Enable> ON | OFF
TRACe:REMote:MODE:FILE:FILTer <Input>, <Output>, <Error>, <Trigger>,
<DeviceClear>, <StatusRegister>, <Connection>, <RemoteLocalEvents>,
<Locking>
Defines the content of the remote trace file.
Parameters:
<Input> ON | OFF
Input data
<Output> ON | OFF
Output data
Remote trace