Measurements
R&S
®
RTE
395User Manual 1326.1032.02 ─ 20
Remote command:
MEASurement<m>:ONViolation:BEEP on page 1306
Stop acq
Stops data acquisition on violation.
Remote command:
MEASurement<m>:ONViolation:ACQStop on page 1306
Print
Prints a screenshot including the measurement results to the printer defined in the
"Print" dialog box (see Chapter 12.4.2, "Printing screenshots", on page 506).
Remote command:
MEASurement<m>:ONViolation:PRINt on page 1307
Save Wfm
Saves the waveform data to the file.
Saves the waveform data to the file specified in [FILE] > "Save" > "Waveform".
Remote command:
MEASurement<m>:ONViolation:WFMSave on page 1307
Trigger Out Pulse
Creates a pulse on the TRIGGER OUTPUT connector on limit violation.
When "Trigger Out Pulse" is used, the trigger control option "Enable trigger out" is dis-
abled. Thus, the trigger-out pulse is created only on limit violation but not when a trig-
ger occurs. The pulse is provided always with the minimum delay of 800 ns, the
"Delay" cannot be set.
See also: "Trigger out signal setup" on page 270.
Remote command:
MEASurement<m>:ONViolation:TRIGgerout on page 1308
Report
Creates and saves a report using the settings defined in "File" menu > "Report Setup".
Remote command:
MEASurement<m>:ONViolation:REPort on page 1307
Start Executable
Starts an external application. Tap "Config Executable" to set the application path and
parameters.
See: Chapter 4.5, "External application", on page 157.
Remote command:
MEASurement<m>:ONViolation:RUNexec on page 1308
Automatic measurements