Measurements
R&S
®
RTE
397User Manual 1326.1032.02 ─ 20
4. Tap the "Main measurement" button.
5. Select the measurement that you want to use for long-term and histogram analysis.
6. Tap "Activate". Select a measurement that you want to add to the quick measure-
ment.
For details on the available measurements, see Chapter 8.2.5.1, "Overview of
amplitude/time measurements", on page 349.
7. Repeat the previous step until the setup is complete.
8. Tap each measurement in the list. If additional settings are needed, adjust them
under "Settings".
9. Tap "Set as QuickMeas" to save the configuration.
The saved configuration remains until you save another QuickMeas setup, there is
no reset.
Set as QuickMeas
Saves the current QuickMeas setup. The saved QuickMeas configuration is used when
you start a new quick measurement. It remains until you save another QuickMeas
setup, or until you reset the instrument to "Factory defaults". The "Set as QuickMeas"
is not available if the current configuration already has been saved.
Quick measurements