"Condition"
Defines the operator to set a specific data, e.g. "Equal" or "Not
Equal") or a range.
"Data Min/Data"
Defines the bit pattern of the data pattern.
In binary format, use the following characters: 1; 0; or X (do not care).
To enter the pattern in other formats use the bit pattern editor, see
Chapter 13.1.5, "Bit pattern editor", on page 521.
"Data Max"
The second data pattern is required to specify a range with conditions
"In range" and "Out of range".
Remote command:
SEARch:TRIGger:HBTO:DATA:DCONdition on page 1694
SEARch:TRIGger:HBTO:DATA:DMIN on page 1694
SEARch:TRIGger:HBTO:DATA:DMAX on page 1695
Index
Sets the specified value or range within this series of data that is considered for the
search. The index setup consists of the condition and one or two index values.
"Condition"
Defines the operator to set a specific index, e.g. "Equal" or a range.
"Index Min/Index"
Defines the bit pattern of the index pattern.
In binary format, use the following characters: 1; 0; or X (do not care).
To enter the pattern in other formats use the bit pattern editor, see
Chapter 13.1.5, "Bit pattern editor", on page 521.
"Index Max"
The second index pattern is required to specify a range with condi-
tions "In range".
Remote command:
SEARch:TRIGger:HBTO:DATA:ICONdition on page 1695
SEARch:TRIGger:HBTO:DATA:IMIN on page 1695
SEARch:TRIGger:HBTO:DATA:IMAX on page 1696
Errors
Sets the type of error events to be searched for.
Preamble Error ← Errors
Searches for a frame with invalid preamble.
Remote command:
SEARch:TRIGger:HBTO:ERRor:PREamble on page 1696
CRC Error ← Errors
Searches for a frame that has a mismatch of the Cyclic Redundancy Check (CRC)
value between the transmitting and receiving device.
Remote command:
SEARch:TRIGger:HBTO:ERRor:CRC on page 1696
SFD Error ← Errors
Searches for a frame with invalid Start Frame Delimiter (SFD).