Protocol analysis
R&S
®
RTE
792User Manual 1326.1032.02 ─ 20
Frame type ← Advanced
Selects the frame type for the custom bus trigger analysis.
You can define individual checking parameters for the fields listed in the "Trigger type
dependent settings".
To specify these parameters, select a field from this list and define the data and/or
index operators and values, or the bit state.
The trigger instant is the last criterion that is fulfilled.
Remote command:
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:ENABle on page 1750
Field name / Condition / Enable ← Advanced
This table lists the field numbers and names in the selected frame together with a sum-
mary of the user settings of checking conditions for each field and a checkbox to
enable the checking.
Select a field in the table to specify the checking conditions for this field in the "Data",
"BitState" and/or "Index" dialog (whichever applies). The condition is only applied, and
the "Condition" is only shown in the table, if "Enable" is checked.
Remote command:
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:ENABle on page 1750
TRIGger<m>:CMSB:ADVanced:FIENable on page 1750
BitState ← Advanced
Defines the bit state to be checked for the selected field. Permitted bit states are "1",
"0" or "X" (do not care).
Remote command:
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:BIT on page 1751
TRIGger<m>:CMSB:ADVanced:BIT on page 1751
Data ← Advanced
Defines for the selected field, how a data check is executed.
"Condition"
Defining specific data or a data range requires to set the operator to
one of the following conditions: equal, not equal, less than, less than
or equal, greater than, greater than or equal, in range, out of range.
"Min"
Specifies data or sets the start value of a data range.
"Max"
Sets the end value of a data range, if "Condition" is set to "In range"
or "Out of range".
Remote command:
TRIGger<m>:CMSB:ADVanced:DMAX on page 1751
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:DMAX on page 1751
TRIGger<m>:CMSB:ADVanced:DMIN on page 1752
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:DMIN on page 1752
TRIGger<m>:CMSB:ADVanced:DOPerator on page 1752
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:DOPerator on page 1752
Index ← Advanced
Defines for the selected field, how an index check is executed.
Custom: Manchester / NRZ (option R&S
RTE-K50)M