Power analysis (option R&S RTE-K31)
R&S
®
RTE
968User Manual 1326.1032.02 ─ 20
To measure and display the power quality, the instrument uses the following measure-
ments and waveforms:
"Meas 6" to measure the voltage
"Meas 7" to measure the current
"Meas 8" to measure the power
"M2" "Math 2" to calculate the power
The used resources are listed in the "Details" tab. See also: Chapter 16.2.3, "Details
tab", on page 966.
Voltage and current results
The voltage and current results are defined as follows:
Result Description
RMS Square root of the mean of the square of the current
or voltage averaged over N cycles
Peak Highest measured magnitude value of the voltage or
current
Crest factor Peak value / RMS value
Frequency Frequency of the signal
Power results
The power in a system is described by several physical quantities: real power, reactive
power, complex power, and phase angle. In Figure 16-1, you can see how these quan-
tities are related if the voltage and the current are sinusoidal signals.