Getting started
R&S
®
RTE
97User Manual 1326.1032.02 ─ 20
1. Press the [SEARCH] key on the front panel (ANALYSIS area).
2. Select "C1" as "Source".
3. Select the search criteria: Enable "Window".
4. In the "Window" tab below, define the search conditions:
a) In the "Upper level" field, enter the result of the "High" level measurement.
b) In the "Lower level" field, enter the result of the "Low" level measurement.
c) As "Vertical condition", select "Exit" to find values that are outside the range
defined by the high and low levels.
5. Select "Enable" to start the continuous search on the acquired data.
6. Close the "Search" dialog box.
7. In the "Search Results" box, select "Show search zoom windows".
The acquisition stops, and the detected overshoots of the last acquisition are listed
in the search result table. The search zoom window shows the last result that was
found. Vertical lines indicate the time values for which a result was found.