Remote control commands
R&S
®
RTE
1145User Manual 1326.1032.02 ─ 20
Parameters:
<Bandwidth> Range: 5000 to 1E+6
Increment: 5000
*RST: 1E+6
Default unit: Hz
Usage: Asynchronous command
ZVC:TYPE <DecimationMode>
Sets the decimation mode for the R&S RT-ZVC probe. Decimation reduces the data
stream of the ADC to a stream of waveform points with lower sample rate and a less
precise time resolution.
Parameters:
<DecimationMode> SAMPle | PDETect | HRESolution
*RST: SAMPle
Usage: Asynchronous command
ZVC:Z<m>:I<n>:BANDwidth?
Queries the bandwidth of the current channel. You can set the probe bandwidth with
ZVC:BANDwidth.
The bandwidth of some current channels is restricted to 300KHz due to their vertical
settings.
Suffix:
<m>
.
1..2
<n> 1..4
Return values:
<Bandwidth> Range: 5000 to 1E+6
Increment: 5000
*RST: 1E+6
Default unit: Hz
Usage: Query only
Asynchronous command
ZVC:Z<m>:I<n>:IMPedance <MeasImp>
Sets the current impedance of the probe channel for power calculations and measure-
ments.
Suffix:
<m>
.
1..2
<n> 1..4
Acquisition and setup