Remote control commands
R&S
®
RTE
1170User Manual 1326.1032.02 ─ 20
TRIGger<m>:ANEDge:CUToff:LOWPass <AnalogCutOffLP>
Frequencies higher than the "Cut-off" frequency are rejected, lower frequencies pass
the filter.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<AnalogCutOffLP> KHZ5 | KHZ50 | MHZ50
KHZ5
5 kHz
KHZ50
50 kHz
MHZ50
50 MHz
*RST: KHZ50
Usage: Asynchronous command
TRIGger<m>:ANEDge:GND <Ground>
Connects the analog signal to the ground.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<Ground> ON | OFF
*RST: OFF
Usage: Asynchronous command
TRIGger<m>:ANEDge:SLOPe <Slope>
Sets the edge for the trigger event.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<Slope> POSitive | NEGative
See Chapter 18.4.3, "Slope parameter", on page 1075.
*RST: POSitive
Usage: Asynchronous command